Ares-G2 (ARES)
Guarantor:
Petr Lepcio, Ph.D.
Instrument status:
Operational
Equipment placement:
CEITEC Nano - B1.15
Ares-G2 by TA Instruments (USA) is the most advanced separated-motor-and-transducer (SMT) rotational rheometer for research and material development. Its separate motor and transducer technology insures the most reliable rheological data through independent control of torque and measurement of stress. It remains the only commercially available rheometer with a dedicated actuator for deformation control, Torque Rebalance Transducer (TRT), and Force Rebalance Transducer (FRT) for independent shear stress and normal stress measurements. An amount of a sample depends on used geometry dimensions but it is typically few milliliters (grams) of measured material.
Photogallery
Specification
- Minimum Transducer Torque in Oscillation 0.05 µN.m
- Minimum Transducer Torque in Steady Shear 0.1 µN.m
- Maximum Transducer Torque 200 mN.m
- Transducer Torque Resolution 1 nN.m
- Transducer Normal/Axial Force Range 0.001 to 20 N
- Maximum Motor Torque 800 mN.m
- Strain Resolution 0.04 µrad
- Min. Angular Displacementin Oscillation 1 µrad
- Max. Angular Displacement in Steady Shear Unlimited
- Angular Velocity Range 1 x 10-6 rad/s to 300 rad/s
- Angular Frequency Range 1 x 10-7 rad/s to 628 rad/s
- Step Change in Velocity 5 ms
- Step Change in Strain 10 ms
- Minimum Transducer Force in Oscillation in axial direction 0.001 N
- Maximum Transducer Force in axial direction 20 N
- Minimum Displacement in Oscillation in axial direction 0.5 µm
- Maximum Displacement in Oscillation in axial direction 50 µm
- Displacement Resolution in DMA mode 10 nm
- Axial Frequency range in DMA mode 1 x 10-5 Hz to 16 Hz
- Temperature Control
- Forced Convection Oven
- Temperature Range -150 to 600 °C
- Advanced Peltier System, APS -10 °C to 150 °C
- FCO Camera Viewer available
Documents